StatSpin Express 4 Horizontal Centrifuge Model # M510. Intended Use: Rapid Separation of plasma and serum from primary g

FDA device recall Z-0807-2010 · posted 2010-03-04 · Terminated

Recall details

Recalling firm
StatSpin, Inc., dba Iris Sample Processing
Reason for recall
Centrifuge shield micro-switch failed due to damage, and the unit opened, ejecting pieces.
Action taken
Iris Sample Processing notified distributors by an Urgent Device Recall letter advising of the recall and the defect. Users are to immediately quarantine the affected products and choose between the two options listed in regards to correcting the affected centrifuge. Customers may return the unit or an upgrade kit with instructions will be sent to upgrade the unit. Questions should be directed to Gail Magrath at 800-782-8774 ext. 6123.
Root cause
Device Design
Status
Terminated
Product code
JQC
Quantity affected
1510 units
Distribution
Worldwide Distribution -- USA, CANADA, AUSTRALIA, JORDAN, UK, BELGIUM, FRANCE, CZECH, RUSSIA, KUWAIT', NORWAY, GREECE, TURKEY, DENMARK, FINLAND, SEOUL, KOREA, and HONG KONG.
Date initiated
2010-01-27
Date posted
2010-03-04
Date terminated
2019-11-13
Location
Westwood, MA