Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel col
FDA device recall Z-2169-2011 · posted 2011-05-10 · Terminated
Recall details
- Recalling firm
- StatSpin, Inc., dba Iris Sample Processing
- Reason for recall
- Rotors may fail due to wire spokes breaking
- Action taken
- Iris notified customers by letter in August 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were requested to install the new rotors and discard the older rotors with the wires. There was no need for customers to return the rotors, but the firm requested customers fax or e-mail the enclosed form to verify that they have installed the new rotors. For any questions customers should call 1-800-782-8774.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JQC
- Quantity affected
- 2090 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, Australia Belgium, Czech Republic, Finland, France, Greece, Kuwait, Norway, Turkey,UK, and Norway.
- Date initiated
- 2010-08-18
- Date posted
- 2011-05-10
- Date terminated
- 2020-09-01
- Location
- Westwood, MA