FDA device recall Z-2234-2008 · posted 2008-08-23 · Terminated

Recall details

Recalling firm
StatSpin, Inc., dba Iris Sample Processing
Reason for recall
EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes.
Action taken
Iris Sample Processing notified customers by telephone followed by an Urgent: Product Recall letter dated 6/19/08. Users are requested to return inventory, to not use product from Lot Number 7296, and to repeat tests where the recalled tubes were used. The firm identified an additional lot : 8031 was effected and recalled the lot to two consignees.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
JKA
Quantity affected
171 vials
Distribution
Worldwide Distribution --- USA including states of CA, CO, FL, KY, KS, GA, NC, NJ, OR, PA, and countries of Canada, Hong Kong, UK.
Date initiated
2008-06-20
Date posted
2008-08-23
Date terminated
2012-05-30
Location
Westwood, MA