StatLock CV Plus w/Pigtail, Catheter Stabilization Device Sterile EO Single Use Only, C.R. Bard, Inc. The StatLock devic

FDA device recall Z-1618-2013 · posted 2013-06-28 · Terminated

Recall details

Recalling firm
Bard Access Systems
Reason for recall
BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized skin irritation. Missing directional arrow may lead to mis-orientation of the StatLock stabilization which may impact the securement of the cathet
Action taken
On 6/18/13, an "Urgent Kit Recall" notification letter was mailed to all customers informing them of the recall and instructing the consignee to remove and return any affected product lot numbers to Bard Access Systems through return goods authorization. Customers may contact 800-290-1689 if questions.
Root cause
Packaging
Status
Terminated
Product code
KMK
Quantity affected
3,875 units.
Distribution
National and Global distribution (Europe, European Economic Area, Argentina, China, and Brazil).
Date initiated
2013-06-10
Date posted
2013-06-28
Date terminated
2014-05-21
Location
Salt Lake City, UT