Atlas FDA

Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942

FDA device recall Z-0576-06 · posted 2006-02-28 · Terminated

Recall details

Recalling firm
Nova Biomedical Corporation
Reason for recall
Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs
Action taken
Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.
Root cause
Other
Status
Terminated
Product code
CHL
Quantity affected
498 units
Distribution
Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan,
Date initiated
2006-01-27
Date posted
2006-02-28
Date terminated
2006-10-19
Location
Waltham, MA