Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that con
FDA device recall Z-0124-2015 · posted 2014-10-24 · Terminated
Recall details
- Recalling firm
- Cook Medical Incorporated
- Reason for recall
- MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that the protective caps were dislodged in the pouch. Further investigation disclosed potential dislodging of protective caps on the needle tips on other distributed products.
- Action taken
- On 9/26/2014, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and institutions Risk Managers/Recall Administration with instructions for returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call Customer Relations at 1-800-457-4500 or 1-812-339-2235.
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- KNW
- Quantity affected
- 1239 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of Arizona, California, Colorado, Hawaii, Iowa, Illinois, Massachusetts, Maryland, Michigan, Missouri, New Mexico, Pennsylvania, Texas, Vir
- Date initiated
- 2014-09-26
- Date posted
- 2014-10-24
- Date terminated
- 2016-02-02
- Location
- Bloomington, IN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| CORAMATE/SPIROTOME SYSTEM (K080095) | Medinvents | 2008-04-11 |
| CORAMATE/SPIROTOME SYSTEM (K060384) | Medinvents | 2006-07-13 |