Atlas FDA

Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that con

FDA device recall Z-0124-2015 · posted 2014-10-24 · Terminated

Recall details

Recalling firm
Cook Medical Incorporated
Reason for recall
MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that the protective caps were dislodged in the pouch. Further investigation disclosed potential dislodging of protective caps on the needle tips on other distributed products.
Action taken
On 9/26/2014, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and institutions Risk Managers/Recall Administration with instructions for returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call Customer Relations at 1-800-457-4500 or 1-812-339-2235.
Root cause
Package design/selection
Status
Terminated
Product code
KNW
Quantity affected
1239 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of Arizona, California, Colorado, Hawaii, Iowa, Illinois, Massachusetts, Maryland, Michigan, Missouri, New Mexico, Pennsylvania, Texas, Vir
Date initiated
2014-09-26
Date posted
2014-10-24
Date terminated
2016-02-02
Location
Bloomington, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CORAMATE/SPIROTOME SYSTEM (K080095)Medinvents2008-04-11
CORAMATE/SPIROTOME SYSTEM (K060384)Medinvents2006-07-13