SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-50 Tests IND REF SPMA_S LOT

FDA device recall Z-1448-2015 · posted 2015-04-15 · Terminated

Recall details

Recalling firm
Vitrolife Inc
Reason for recall
The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution was added.
Action taken
FertiPro sent an Urgent: Medical Device Recall letter dated December 22, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to check if they have stock of the concerned lots and provide a confirmation that they have destroyed the affected products. Distributors were asked to inform their customers and instruct them to do the same. For further questions you may call (858) 824-0888.
Root cause
Component design/selection
Status
Terminated
Product code
KTO
Quantity affected
45 kits
Distribution
US Distribution to the state of : OR., and Internationally to Argentina.
Date initiated
2014-12-22
Date posted
2015-04-15
Date terminated
2015-05-05
Location
San Diego, CA