Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in
FDA device recall Z-2863-2017 · posted 2017-07-14 · Terminated
Recall details
- Recalling firm
- Spectranetics Corporation
- Reason for recall
- Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device is in use in an off-label manner.
- Action taken
- Spectranetics sent a Field Safety Notice letter via certified mail on June 7, 2017, to all customers who have ordered the affected model numbers since 23May2016 reiterating the risks associated with off-label use of the device (i.e., not activating the laser within a stent and not until all contrast media is flushed from the treatment area). The product continues to meet specifications and does not need to be returned. Customers with questions were encouraged to call their local Spectranetics Sales Representative or Spectranetics Customer Service Department. Customer Service can be reached by phone at 800-231-0978 (Option 2), by fax at 877-447-2022, or by email at customer.service@spnc.com. Customer Service is available Monday thru Friday, From 7AM-5PM MST.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MCW
- Distribution
- Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, Hawaii and
- Date initiated
- 2017-06-06
- Date posted
- 2017-07-14
- Date terminated
- 2021-07-08
- Location
- Colorado Springs, CO