Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bags with an attached in

FDA device recall Z-2314-2008 · posted 2008-09-20 · Terminated

Recall details

Recalling firm
Spacelabs Healthcare, Incorporated
Reason for recall
Intermittent contact of temperature probe adaptors can cause a loss of temperature monitoring which can cause a delay in care when an alarm is missed.
Action taken
Starting July 3, 2008 the firm issued a "MEDICAL DEVICE CORRECTION" letter notifying their customers of the potential hazards with the use of probe adapter, PN 700-031-00. All letters have a return receipt to verify each customer received the medical device correction letter. The firm will replace all PN 700-0031-00 adaptors with a replacement adaptor from their initial distributor. The firm initiated sending new adaptors separately, by mail, with a copy of the same Medical Device Correction letter to the firm's customers. The firm has requested their customers to throw away any adapter that is being recalled. Technical Support will be available for customer questions and technical questions and to ensure each unit is upgraded. Contact Spacelabs Healthcare at 425-657-7200 ext 5970 if you have questions.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DSI
Quantity affected
3,733
Distribution
Nationwide and Worldwide to medical facilities.
Date initiated
2008-07-02
Date posted
2008-09-20
Date terminated
2011-06-07
Location
Issaquah, WA