Atlas FDA

Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-09 transmitter, 9

FDA device recall Z-1202-06 · posted 2006-07-06 · Terminated

Recall details

Recalling firm
Spacelabs Medical Incorporated
Reason for recall
Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
Action taken
On May 5, 2006 a letter was issued to all consignees, via return receipt requested. The letter advised consignees of the issue and provided some ways to mitigate the risk. Anyone experiencing the anomaly is to document the failure and report it to the firm. A field service representative will visit each consignee to optimize the antenna system to reduce intermodulations. After determining the root cause and validating a fix, the firm issued a second letter dated August 21, 2006 to their consignees. The letter advised consignees of three issues that needed 'improvement' and stated a service engieneer will schedule a time to visit and install the improvements.
Root cause
Other
Status
Terminated
Product code
DSI
Quantity affected
22 telemetry systems; 1229 transmitters
Distribution
The systems were distributed to 22 hospitals/medical centers nationwide and to 2 Spacelabs personnel. There is no international distribution.
Date initiated
2006-05-05
Date posted
2006-07-06
Date terminated
2010-12-27
Location
Issaquah, WA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODEL 90478 & 90479 TELEMETRY REC. MODULE&HOUSING (K925510)Spacelabs Medical, Inc.1994-10-27