Atlas FDA

Sorin Stockert Heater-Cooler 3T, 200 V / 50Hz / 60 Hz Temperature control for extracorporeal perfusion of durations up t

FDA device recall Z-2081-2015 · posted 2015-07-15 · Terminated

Recall details

Recalling firm
Sorin Group USA, Inc.
Reason for recall
Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Action taken
Sorin Group issued a Field Safety Notice dated June 15, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and identify any affected devices. For each unit customers were instructed to determine if the device has been maintained according to the Instructions for Use. If yes, customers should strictly adhere to the new Instructions for Use. Customers were also provide with a Response form to confirm they received, read and understood the Field Notice. Customers were instructed to return the completed form to assist in monitoring the effectiveness of the communication. For technical support customers should call 1-800-221-7943, ext 6355. For questions regarding this recall call 303-467-6306.
Root cause
Error in labeling
Status
Terminated
Product code
DWC
Quantity affected
131
Distribution
Worldwide Distribution - US (nationwide) and Internationally to AL, AT, BE, BG, CH, CY, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HR, HU, IE, IT, LT, LU, LV, MC, MK, MT, NL, NO, PL, PT, RO, RS, SE, SK, AE,
Date initiated
2015-06-15
Date posted
2015-07-15
Date terminated
2018-07-25
Location
Arvada, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STOCKERT HEATER-COOLER SYSTEM 3T (K052601)Sorin Group Deutschland GmbH2006-06-06