Somatom Sensation 64 X-Ray System, Model number 8377520. This is a computed tomography x-ray system which is intended to

FDA device recall Z-0321-2009 · posted 2009-01-05 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
software error - does not allow protocol to properly synchronize with ECG triggers
Action taken
The recalling firm issued a Customer Safety Advisory Notice dated 9/3/08 to affected customers via Update Instructions CT045/08/S. This letter informs customers of the potential issued and provides instruction to avoid its occurrence. For additional information, contact Siemens Medical Solutions USA, Inc. at 1-610-448-4500.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
9 units
Distribution
Nationwide
Date initiated
2008-09-03
Date posted
2009-01-05
Date terminated
2009-09-23
Location
Malvern, PA