SOMATOM Emotion 6. Computed tomography X-ray system

FDA device recall Z-2452-2008 · posted 2008-09-21 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.
Action taken
The firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JAK
Quantity affected
53 units
Distribution
Nationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.
Date initiated
2008-08-05
Date posted
2008-09-21
Date terminated
2008-11-07
Location
Malvern, PA