Atlas FDA

SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data fro

FDA device recall Z-1933-2015 · posted 2015-06-27 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Action taken
A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
Root cause
Device Design
Status
Terminated
Product code
JAK
Quantity affected
40 total
Distribution
Nationwide Distribution.
Date initiated
2015-06-04
Date posted
2015-06-27
Date terminated
2017-03-24
Location
Malvern, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SOMATOM EMOTION 6 CT SYSTEMS, SOMATOM EMOTION 16 (K133424)Siemens Medical Solutions USA, Inc.2014-01-17