Somatom Definition Model AS/AS+, Model number 8098027. Intended use: Computed Tomography X-ray system

FDA device recall Z-1675-2009 · posted 2009-07-30 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Support mechanism may not engage properly due to improper stowage after opening which may result in internal component damage or potential injury.
Action taken
The firm has issued a modification of the telescope support arm parking bracket via Update Instruction CT028/09/S. A firm service representative will install the modified support arm parking brackets on the affected systems.
Root cause
Equipment maintenance
Status
Terminated
Product code
JAK
Quantity affected
93 units
Distribution
Nationwide Distribution -- including states of AR, AZ, CA, CO, CT, FL, IA, IL, IN, KS, KY, MA, MD, MN, MO, MT, NC, ND, NE, NH, NY, OH, PA, SC, SD, TN, VA, WA, WI, and WY.
Date initiated
2009-05-29
Date posted
2009-07-30
Date terminated
2009-09-01
Location
Malvern, PA