SOMATOM Definition AS, Computed tomography system Product Usage: SOMATOM Definition Flash systems are intended to produc

FDA device recall Z-0604-2017 · posted 2016-11-17 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.
Action taken
Siemens' sent a Customer Safety Advisory Notice letter dated October 19, 2016 to affected customers. The letter identified the affected product, problem, potential risks, how to avoid the potential risk related to this issue and actions to be taken.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JAK
Quantity affected
762 systems distrtibuted in U.S.
Distribution
US Nationwide Distribution
Date initiated
2016-10-19
Date posted
2016-11-17
Date terminated
2017-03-13
Location
Malvern, PA