SOMATOM CT System Emotion 16. Computed tomography x-ray system.

FDA device recall Z-0793-06 · posted 2006-04-26 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
When using the LungCare feature in the SOMATOM CT Systems some marks made on CT images may not be saved.
Action taken
The recalling firm has issued a software patch to the affected customers per Update Instructions CT086/05/S and CT087/05/S. By letter 3/9/2006. In addition, a Siemens Service Engineer will visit affected sites to install this software patch which will resolve this issue.
Root cause
Other
Status
Terminated
Product code
JAK
Quantity affected
5 units
Distribution
Nationwide. The product was shipped to AK, AZ, CA, DE, FL, GA, IL, IN, KY, LA, MD, ME, MI, MT, NC, NY, OH, OK, PA, PR, TX, UT, VA, WA and WI.
Date initiated
2006-03-09
Date posted
2006-04-26
Date terminated
2006-07-19
Location
Malvern, PA