Atlas FDA

SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX1

FDA device recall Z-0748-2024 · posted 2024-01-18 · Open, Classified

Recall details

Recalling firm
Olympus Corporation of the Americas
Reason for recall
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
Action taken
On December 1, 2023, the firm issued URGENT MEDICAL DEVICE CORRECTION letters to affected customers. Customers were provided with copies of the labeling addendum to add to remaining inventory. Customers may continue to use the product, but this should be done in accordance with the labeling addendum. If you require additional information, please contact Olympus at (647) 999-3203 or at Cynthia.Ow@Olympus.com.
Root cause
Labeling Change Control
Status
Open, Classified
Product code
GEX
Quantity affected
82,200 total units (5 per box)
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.
Date initiated
2023-12-01
Date posted
2024-01-18
Location
Center Valley, PA