Solara CRT-P MRI, Model Numbers: a) W1TR03, b) W1TR06, c) W4TR03, d) W4TR06

FDA device recall Z-2513-2020 · posted 2020-07-10 · Open, Classified

Recall details

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Reason for recall
There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
Action taken
The firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions.
Root cause
Software design
Status
Open, Classified
Product code
NVZ
Quantity affected
16667 devices
Distribution
worldwide
Date initiated
2020-06-04
Date posted
2020-07-10
Location
Mounds View, MN