Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI

FDA device recall Z-2669-2026 · posted 2026-07-02 · Open, Classified

Recall details

Recalling firm
Laborie Medical Technologies Corp
Reason for recall
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Action taken
Laborie Medical Technologies notified consignees on about 05/21/2026 via letter. Consignees were instructed to always use the pre-check instructions provided in the IFU to ensure that the device is operating properly, ensure all operators are aware of the outlined IFU updates in the customer notification to include the following: 1. Updated 3-Minute Blow-Out Moisture Removal Maintenance Procedure (New IFU Section 3.0 and provided in the attached insert). 2. Laborie has added the following to the revised IFU in addition to the blow-out procedure presented in Attachment 1: " A Device Use Life Statement in Section 1.8. " The statement in Section 2.4.3. Consignees were also instructed to review the Enclosed IFU Insert Section 3.0 for the updated Blow-Out Moisture Removal Maintenance Procedure and ensure all clinical users are trained to the new maintenance procedure. Laborie recommends that customers print-out the Blow-Out procedure and post it on or near the device for awareness of the change in the Blow-Out Cycle. Distribute this notification and IFU insert to all clinical operators, supervisors and trainers, biomedical engineering / technical support staff, and any personnel responsible for system cleaning, setup, or maintenance.
Root cause
Device Design
Status
Open, Classified
Product code
FFX
Quantity affected
372 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germ
Date initiated
2026-05-21
Date posted
2026-07-02
Location
Portsmouth, NH