SoftWeb version 4.0.4.10.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Date: 05/16/2006.
FDA device recall Z-0099-2011 · posted 2010-10-20 · Terminated
Recall details
- Recalling firm
- SCC Soft Computer
- Reason for recall
- If a case number is re-used in SoftWeb with SoftPath, the SoftPath case is linked to the original MRN in SoftWeb. This causes SoftPath cases being linked to the wrong patient in SoftWeb. Incorrect results on the patient with original MRN and no results on the patient with the second MRN (and the reused case number) are displayed in SoftWeb Results Query. SCC Soft Computer issued a Correction for
- Action taken
- SCC Soft Computer issued a Risk-to-Health notification with Client Notification Verbiage to the affected client using SCC Soft Computer's proprietary Task Management System on June 8, 2006. SCC recommended SoftPath/SoftWeb users NOT re-use SoftPath case numbers when cases are cancelled. The generic option setting "Specimen Registration - Standard Settings - Minimum Time to Reactivate Case Number" should be set to "0" so the user is not prompted to reuse the same case number. Client was asked to acknowledge receipt of the task and grant permission to load the required software correction.
- Root cause
- Software design
- Status
- Terminated
- Product code
- LNX
- Quantity affected
- 1
- Distribution
- Nationwide Distribution: 1 hospital in Alaska.
- Date initiated
- 2006-06-08
- Date posted
- 2010-10-20
- Date terminated
- 2010-11-03
- Location
- Clearwater, FL