Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medic
FDA device recall Z-1181-2024 · posted 2024-02-22 · Open, Classified
Recall details
- Recalling firm
- PTW-FREIBURG
- Reason for recall
- Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
- Action taken
- PTW Dosimetry issued Urgent Medical Device Correction letter via email on 2/8/24. Letter states reason for recall, health risk and action to take: PTW recommends not using the Suppress gamma calculation function for ROIs where a high proportion of voxels are to be suppressed in the gamma calculation. Since the malfunction is limited to the GPR calculation of ROIs and does not affect the GPR calculation of the VERIQA Total volume, it is recommended to rely on the comparison based on VERIQA Total. PTW will correct the malfunction in the next software version. The corrected version will be made available to all VERIQA users. The above information should be taken into account by the customer until the installed VERIQA version has been updated. Pass this notice on to everyone at your facility/hospital or at another facility/hospital where the equipment is located who needs to know about it.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- IYE
- Quantity affected
- 1 US; 65 worldwide
- Distribution
- GA
- Date initiated
- 2024-02-08
- Date posted
- 2024-02-22
- Location
- Freiburg Im Breisgau, Germany
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VERIQA (K213370) | Ptw-Freiburg Physikalisch-Technische-Werkstaetten | 2022-03-30 |