Software: SA BASE 3.1.4.x, 3.1.6.x and SA LabMic 4.0.0.x, 4.0.1.x, 4.0.2.x, 4.0.3.x, 4.0.4.x. SoftLab is a laboratory in
FDA device recall Z-0261-2012 · posted 2011-11-21 · Terminated
Recall details
- Recalling firm
- SCC Soft Computer
- Reason for recall
- On 08/08/2011 SCC Soft Computer, Clearwater, FL initiated a correction on following SoftLab used with: SA BASE 3.1.4.x, 3.1.6.x and SA LabMic 4.0.0.x, 4.0.1.x, 4.0.2.x, 4.0.3.x, 4.0.4.x. A client reported a flagging error where the result was flagged as critical in the Instrument Menu but the printed report did not show the corresponding flag. All affected clients were notified of the issue on 0
- Action taken
- SCC Soft Computer notified all affected customers on August 8, 2011, of the recall via the SCC Soft Computer proprietary Task management system to manage client communication. Clients use this tool to log in requests for services and receive and provide updates to any issue. Each client user has a unique password and every action is date and time-stamped. Each Task has a unique identifier and a full audit trail. SCC recommended to the customer until Live on the correction avoid making edits in the instrument menu. For further questions please call (727) 789-0100.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 255
- Distribution
- Worldwide Distribution - USA ( nationwide) and Canada.
- Date initiated
- 2011-08-08
- Date posted
- 2011-11-21
- Date terminated
- 2013-11-22
- Location
- Clearwater, FL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.