SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to be used in a

FDA device recall Z-0363-2012 · posted 2011-12-02 · Terminated

Recall details

Recalling firm
SCC Soft Computer
Reason for recall
On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were incorrect. The HPV results were from the previous order on the patient and not the current order. All affected clients were notified about the issue on 07/20/2011.
Action taken
SCC Soft Computer sent a "CORRECTION COMMUNICATION" dated July 20, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. SCC personnel will review the customized Lab Results AP macro and/or the Micro Results AP macro on customer systems and will either discuss with them the irregularities requiring change or reconfigure macros as necessary.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
97
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Date initiated
2011-07-20
Date posted
2011-12-02
Date terminated
2013-05-21
Location
Clearwater, FL