SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC
FDA device recall Z-0088-2011 · posted 2010-10-19 · Terminated
Recall details
- Recalling firm
- SCC Soft Computer
- Reason for recall
- Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for SoftMic Versions 4.0.4.7-4.5.1.3 software due to this potential hazard.
- Action taken
- SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software. Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used. Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.
- Root cause
- Software design
- Status
- Terminated
- Product code
- LNX
- Quantity affected
- 35
- Distribution
- Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.
- Date initiated
- 2010-07-19
- Date posted
- 2010-10-19
- Date terminated
- 2012-07-16
- Location
- Clearwater, FL