SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medica

FDA device recall Z-0466-2012 · posted 2012-01-11 · Terminated

Recall details

Recalling firm
SCC Soft Computer
Reason for recall
On 08/25/2011, SCC Soft Computer initiated a correction on the following SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. For clients using SoftMic Result Entry, when a certain series of keystrokes on the test grid are performed while entering results and statusing a test, the test status may not be maintained upon saving. The status may be removed or copied from another test on the order
Action taken
SCC Soft Computer initiated a Field Correction dated August 25, 2011. to all affected customers. Customers with the affected software have been notified and are scheduling receipt of the hot fix.. Further questions call (727) 789-0100.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
144
Distribution
Worldwide Distribution -- USA (nationwide) and Canada.
Date initiated
2011-08-25
Date posted
2012-01-11
Date terminated
2014-10-03
Location
Clearwater, FL