SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medica
FDA device recall Z-0466-2012 · posted 2012-01-11 · Terminated
Recall details
- Recalling firm
- SCC Soft Computer
- Reason for recall
- On 08/25/2011, SCC Soft Computer initiated a correction on the following SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. For clients using SoftMic Result Entry, when a certain series of keystrokes on the test grid are performed while entering results and statusing a test, the test status may not be maintained upon saving. The status may be removed or copied from another test on the order
- Action taken
- SCC Soft Computer initiated a Field Correction dated August 25, 2011. to all affected customers. Customers with the affected software have been notified and are scheduling receipt of the hot fix.. Further questions call (727) 789-0100.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 144
- Distribution
- Worldwide Distribution -- USA (nationwide) and Canada.
- Date initiated
- 2011-08-25
- Date posted
- 2012-01-11
- Date terminated
- 2014-10-03
- Location
- Clearwater, FL