SoftLab with SA INST versions: 3.1.6.12, 4.0.1.26-4.0.1.32, 4.0.2.21-4.0.2.51, and 4.0.3.5-4.0.3.13. Product UsageSoftLa

FDA device recall Z-2143-2012 · posted 2012-08-03 · Terminated

Recall details

Recalling firm
SCC Soft Computer
Reason for recall
On 10/17/2011 SCC Soft Computer, Clearwater, FL initiated a correction on SoftLab with SA INST versions: 3.1.6.12, 4.0.1.26-4.0.1.32, 4.0.2.21-4.0.2.51, and 4.0.3.5-4.0.3.13. A client reported an issue which caused two different sets of results from the Sysmex series of Hematology instruments to display under the same order on the instrument menu.
Action taken
SCC Soft Computer's proprietary Task Management System sent a correction communication letter to all consignees with affected software. The letter identified the affected product, problem and corrective actions to be taken. The letter also recommended alternative working solution if the corrective actions cannot be used. Consignees with affected software are scheduling receipt of the software correction. For questions call 727-789-0100.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
88
Distribution
US Nationwide Distribution
Date initiated
2011-10-17
Date posted
2012-08-03
Date terminated
2014-09-03
Location
Clearwater, FL