SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research
FDA device recall Z-1038-2015 · posted 2015-02-02 · Terminated
Recall details
- Recalling firm
- SCC Soft Computer
- Reason for recall
- The interface fails to send abnormal flags for Reference Lab test results.
- Action taken
- The firm sent out 2 CORRECTION COMMUNICATIONS informing consignees of functionality in release of 4.0.7 HIS Add-on interfaces which SCC has designated to be a potential patient Risk-to-Health issue. Further the corrective action indicated HIS releases 4.0.7.2 and higher will be corrected. Hot Fixes have been developed for HIS Add-on patches 4.0.7.0 and 4.0.7.1. Lastly the communication requests that the user "please acknowledge receipt of this task and grant permission to load the mandatory hot fix(es). Thank you."
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 48
- Distribution
- Distributed US (nationwide) in the states of OH, GA, NJ, WI, FL, NH, IL, NY, IN, CA, AL, VA, MO, MI, KY, MA, PA, CO, ME, TN, WA, MN, SD, LA, NC, and NV and the country of Canada.
- Date initiated
- 2014-05-20
- Date posted
- 2015-02-02
- Date terminated
- 2016-05-16
- Location
- Clearwater, FL