SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, F

FDA device recall Z-0344-2011 · posted 2010-11-10 · Terminated

Recall details

Recalling firm
SCC Soft Computer
Reason for recall
When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result.
Action taken
Clients were notified via the proprietary communication software on 02/16/2010. The Notification Verbiage identified the affected product and informed clients of the potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
31
Distribution
Nationwide Distribution - US and Puerto Rico.
Date initiated
2010-02-16
Date posted
2010-11-10
Date terminated
2012-07-17
Location
Clearwater, FL