Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 51

FDA device recall Z-0033-2012 · posted 2011-10-11 · Terminated

Recall details

Recalling firm
Isopure Corp
Reason for recall
Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.
Action taken
Isopure sent a letter dated May 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Isopure devised an immediate solution which must be implemented at their faclity in conjunction with the disinfection procedure. Customers were asked to inform all appropriate personnel, clinical and technical, who may be involved in operating the system during the normal disinfection schedule. On 5/19/2011, the firm sent letters to their customers via US mail and email. They also telephoned their customers to notify them of the problem.
Root cause
Software design
Status
Terminated
Product code
FJP
Quantity affected
11 units
Distribution
Nationwide Distribution including AR, CT, GA, MA, ME, NJ, PA, TN and WV.
Date initiated
2011-05-19
Date posted
2011-10-11
Date terminated
2013-02-19
Location
Simpsonville, KY