Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction

FDA device recall Z-1234-2025 · posted 2025-02-26 · Open, Classified

Recall details

Recalling firm
Scientia Vascular, Inc.
Reason for recall
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
Action taken
On 01/02/2025, the firm initially notified customer via email an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that through internal evaluation of finished goods of the Socrates 38 Aspiration Catheter, one lot of pouches containing the 127 cm Socrates 38 Aspiration Catheter was identified with pouches that had a gap in the sterile barrier seal. This was a result of the pouch not being centered with the sealing bar during the sealing process. On 01/10/2025, The firm sent an updated letter informing customers of additional product lots were added to the Recall. Customers are instructed to: Segregate Recalled Product. Immediately remove all impacted (see list above) Socrates 38 Catheter(s) from their inventory (regardless of its location) and segregate this product in a secure location for return to Scientia Vascular, Inc. In situations where a device has already been used in a patient procedure, closely monitor patients for signs and symptoms of post-surgical infection. Customer Service will be available to provide additional information and provide replacement orders at customercare@scientiavascular.com or (888)385-9016; Monday through Friday, 8:00 AM to 4:30 PM, Mountain Time.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
NRY
Quantity affected
120 devices
Distribution
U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.
Date initiated
2025-01-02
Date posted
2025-02-26
Location
West Valley City, UT