SMV FX-40 Nuclear Camera

FDA device recall Z-1393-04 · posted 2004-08-24 · Terminated

Recall details

Recalling firm
General Electric Med Systems
Reason for recall
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
Action taken
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
Root cause
Other
Status
Terminated
Product code
IYX
Quantity affected
65
Distribution
Inside the US. Manufactured and distributed from 1993 - 2000.
Date initiated
2004-07-30
Date posted
2004-08-24
Date terminated
2006-07-24
Location
Waukesha, WI