Smiths Pneupac babyPAC portable ventilator, model 100. Designed for use by qualified medical caregivers, paramedics and

FDA device recall Z-0436-06 · posted 2006-01-24 · Terminated

Recall details

Recalling firm
Smiths Medical PM, Inc.
Reason for recall
A potential safety related problem associated with the Pnuepac babyPac Ventilator model 100 has been discovered in that the interface between the air input hose and the air input fitting may leak, and if tightened too much, the air input fitting may loosen when trying to remove the air hose.
Action taken
Consignees were provided a Safety Action Bulletin, number 05-SAB05, dated 12/05/05. This requests that affected devices be quarantined, Smiths Medical be contacted and devices ultimately be returned to Smiths Medical Pm, Inc. Service Departmentfor repair.
Root cause
Other
Status
Terminated
Product code
BTL
Quantity affected
44
Distribution
CA, CO, IN,MN, NY, PA, TN, TX
Date initiated
2005-12-05
Date posted
2006-01-24
Date terminated
2006-12-26