Atlas FDA

Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubin

FDA device recall Z-2527-2025 · posted 2025-09-04 · Open, Classified

Recall details

Recalling firm
ICU Medical Inc.
Reason for recall
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Action taken
ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 08/06/2025 via registered mail and email. The notice explained the issue, potential risk can requested the following: Customer Required Actions: Please complete the following actions below. 1. Review your inventory of CADD Medication Cassette Reservoirs to confirm if any of the affected lots are within your possession. 2. Quarantine the affected product and destroy or discard it immediately following your institution s process for destruction or discarding. 3. Return the attached Customer Response Form to ICUMedical8176@sedgwick.com within ten (10) days of receipt to acknowledge your understanding of this notification. 4. DISTRIBUTORS: Immediately forward this notice to all users to whom you distributed these products. Request that they complete the Customer Response Form and return it to ICUMedical8176@sedgwick.com. For further inquires: - To report adverse events or product complaints: Global Complaint Management: productcomplaints@icumed.com; 1-(866)-216-8806 - Additional information or to request replacement product: Customer Service: customerservice@icumed.com; 1-(800)-258-5361 - Questions about this communication: Field Action Processing: ICUMedical8176@sedgwick.com; 1-(844)-861-6220
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
FPA
Quantity affected
15546 units
Distribution
Worldwide distribution.
Date initiated
2025-08-06
Date posted
2025-09-04
Location
Minneapolis, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.