Atlas FDA

Smiths Medical BCI(R) Advisor(R) Vital Signs Monitor, model 9200, catalog numbers: 920654225, 920674225, 920674235, 9254

FDA device recall Z-1326-2009 · posted 2009-05-15 · Terminated

Recall details

Recalling firm
Smiths Medical PM, Inc.
Reason for recall
Recent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket, resulting in the monitor falling off. There have been no adverse events or injuries reported.
Action taken
An Urgent Product Safety Information, Voluntary Recall Notification, dated 12/22/08, was sent to affected consignees. The notification contained a Safety Action Bulletin which provided actions to be taken. The Safety Action Bulletin described the issue, requested customer to contact Smiths Medical to ensure proper mounting hardware is utilized for those consignees using a roll stand or wall mount. If the consignee further distributed the product, they were asked to forward on the Safety Action Bulletin to the end user. Page 3 of the Safety Action Bulletin was asked to be completed and returned to Smiths Medical PM, Inc. A second "Urgent Product safety Information Voluntary Recall Notification" letter dated January 16, 2009 was sent to consignees. The second letter included a risk statement associated with the recalled product.
Root cause
Device Design
Status
Terminated
Product code
CCK
Quantity affected
143 monitors (17 US, 126 OUS)
Distribution
Worldwide Distribution -- USA including states of FL, GA, NJ, NY, and TN and countries of Argentina, Brazil, Ecuador, Greece, Peru, Romania, Russia, Singapore, Slovenia, Trinidad and Tobago, United Ki
Date initiated
2008-12-22
Date posted
2009-05-15
Date terminated
2011-12-17
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ADVISOR, MODEL 9200 (K030407)Bci, Inc.2003-08-19
BCI ADVISOR VITAL SIGNS MONITOR (9200) (K010770)Bci, Inc.2001-06-12
BCI ADVISOR VITAL SIGNS MONITOR (9200) (K011177)Bci, Inc.2001-05-23
VITAL SIGNS MONITOR; ADVISOR MODEL # BCI 9200 (K982279)Bci Intl., Inc.1998-09-25