Smith & Nephew, Inc., Exu-Dry Wound Burn Dressing, Product Number: 5999M24, Exu-Dry Dressing, 15" x 24", Smith & Nephew,
FDA device recall Z-1213-2008 · posted 2008-03-12 · Terminated
Recall details
- Recalling firm
- Smith And Nephew, Inc. Wound Management Division
- Reason for recall
- Sterility (package integrity) compromised: Distribution of product that did not meet specifications; open seals.
- Action taken
- Consignees were notified of recall by certified mail, Urgent Device Recall letter dated 11/21/2007, which was sent to each consignee. The recalled product will be returned to the firm by pre-paid FedEX return labels. A second letter is to be issued to each distributor for recall to the retail level, as the firm's original recall strategy of distributor level recall, was deemed inappropriate by CDRH
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- MCY
- Quantity affected
- 1,710 dressings
- Distribution
- Worldwide-USA, England, Japan, South Africa, New Zealand, Australia and Canada.
- Date initiated
- 2007-11-21
- Date posted
- 2008-03-12
- Date terminated
- 2008-12-15
- Location
- Largo, FL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.