smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur

FDA device recall Z-0281-2025 · posted 2024-10-31 · Open, Classified

Recall details

Recalling firm
Smith & Nephew Inc
Reason for recall
Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.
Action taken
Smith & Nephew issued an "Urgent Medical Device RECALL Notice" to its consignees on 10/02/2024 via email. The notice explained the issue, potential risk of use of the affected product, and requested the return of the affected devices. Return affected product to: Smith+Nephew | Attn: Quality Hold and Field Action Returns Department | 1450 E Brooks Road, Building G | Memphis, TN 38116 Distributors, sales reps, and district offices were directed to notify their customers. If you have any questions or concerns regarding this recall, please contact FieldActions@smith nephew.com.
Root cause
Packaging change control
Status
Open, Classified
Product code
JWH
Quantity affected
5 units
Distribution
US Distribution to states of: MS, TN, NC, CA, GA, DE, AL, and OUS distribution to country of: Canada
Date initiated
2024-10-02
Date posted
2024-10-31
Location
Memphis, TN