Smith & Nephew Aimer Endofemoral 4mm Offset (Red) Catalog Number: 72201715F (Refurbished) ACUFEX Surgical Instruments ar
FDA device recall Z-0047-2011 · posted 2010-10-12 · Terminated
Recall details
- Recalling firm
- Smith & Nephew, Inc. Endoscopy Division
- Reason for recall
- Endofemoral Aimer tip may separate due to failure of the soldered joint
- Action taken
- Smith & Nephew sent a "Medical Device Recall Information" letter dated April 29, 2010, with an attached "Urgent-Product Recall 1st Notification - Urgent" response form via Federal Express on April 29, 2010, to all customers. An e-mail was sent to the contacts in the identified OUS countries alerting them of the recall. ( Note: On June 24, 2010, the firm expanded the recall to include 4 additional catalog numbers 7220l713F, 72201714F, 72201715F, 722017l6F that are refurbished and one additional lot number to catalog Number: 72201714.) The letter described the product, problem and action to be taken by customers. The customers were instructed to immediately check stock for affected product; immediately remove the items from stock and set aside for quarantine; complete and return the attached Urgent-Product Recall 1st Notification - Urgent" response form, even if no product is being returned, via fax at +1-508-261-3636 attention Cindy Burns or email a PDF to Cindy.Burns@smith-nephew.com; call for a return Authorization number at 1-800-343-5717 Option #3, and return product to Smith & Nephew, Inc., Endoscopy Division as soon as possible. If you have any questions or require additional information, please contact Cindy Burns at 1-508-261-3655.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- NBH
- Quantity affected
- 10 units
- Distribution
- Worldwide distribution: USA and countries including: Australia, Austria, Belgium, Brazil, Canada, Costa Rica, Denmark. Dubal, Finland, France. Germany, Greece, Hong Kong, India, Italy, Japan, Malaysia
- Date initiated
- 2010-04-29
- Date posted
- 2010-10-12
- Date terminated
- 2012-05-09
- Location
- Andover, MA