SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid an
FDA device recall Z-2279-2013 · posted 2013-09-26 · Terminated
Recall details
- Recalling firm
- CareFusion 303, Inc.
- Reason for recall
- CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.
- Action taken
- CareFusion sent an Urgent Medical Device Recall Notification dated August 30, 2013, to all affected customers. The letter informed the customers of the problem identified, issues, potential risk, and the immediate actions to be taken. Customers were instructed to immediately check their inventory for the affected product, if any affected product is encountered, customers were instructed not to use. Customers were instructed to complete the customer response form and return the form to CareFusion with the affected product for replacement.. A distributor letter was also sent to distributors on August 30, 2013, to inform them of the problems identified and the actions to be taken. Distributors were instructed to contact the Customer Support Team, return customer response form, and if they have questions then they were instructed to contact the CareFusion Support Center at 1-800-562-6018. For questions regarding this recall call 858-617-4000.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- FPA
- Quantity affected
- 600 units
- Distribution
- CA, FL, AZ, and TN
- Date initiated
- 2013-08-30
- Date posted
- 2013-09-26
- Date terminated
- 2013-10-28
- Location
- San Diego, CA