SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107. Product Usage: The use of the SMA

FDA device recall Z-1691-2015 · posted 2015-05-28 · Terminated

Recall details

Recalling firm
Biosense Webster, Inc.
Reason for recall
Biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an RMT procedure the SMARTABLATE RF Generator inadvertently delivered RF energy.
Action taken
Biosense Webster sent an Urgent Field Safety Notification Letter dated March 18, 2015 via FedEx to all their customers that received the affected SMARTABLATE Generator System. The letter informs the customers of the problems identified and the actions to be taken. Biosense Webster is taking precautionary safety measures of reinforcing the Warnings in the Instructions for Use (IFU) for the SMARTABLATE System Foot Pedal. Customers with questions related to the are instructed to contact their local Biosense Webster sales representative or call (866) 473-7823, Monday through Friday from 7:00AM to 8:00PM EST. Customers with questions related to the voluntary field removal certification form and its return, are instructed to contact Elie Ghattas Field Action Coordinator at (909) 839-7281.
Root cause
Device Design
Status
Terminated
Product code
OAE
Quantity affected
115 units
Distribution
Worldwide Distribution - US Nationwide in the state of: CA, IL, KY, NJ, PA and the countries of: Czech Republic, Denmark, Finland, France, Germany, and Sweden.
Date initiated
2015-03-19
Date posted
2015-05-28
Date terminated
2015-06-17
Location
Irwindale, CA