Skytron Infinity Series surgical light with triple lightheads, 5 bulbs in each of 3 - 22' diameter lightheads, Manufactu

FDA device recall Z-0532-2007 · posted 2007-02-22 · Terminated

Recall details

Recalling firm
Skytron, Div. The KMW Group, Inc
Reason for recall
If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
Action taken
The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Root cause
Other
Status
Terminated
Product code
FTD
Quantity affected
5,000 for all models.
Distribution
Nationwide and Canada.
Date initiated
2007-01-08
Date posted
2007-02-22
Date terminated
2009-11-18
Location
Grand Rapids, MI