Atlas FDA

SJM Pericardial Patch, with EnCap Technology 5x10 cm, Model Number REF C0510; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTF

FDA device recall Z-0532-2026 · posted 2025-11-25 · Open, Classified

Recall details

Recalling firm
Glycar SA Pty., Ltd.
Reason for recall
The impacted lot may not meet the required tensile strength specification.
Action taken
Glycar issued a Stock Withdrawal Notification to its sole distributor on 10/09/2025. The notice explained the issue and requested the following: "Required Actions 1. Immediately identify and segregate all affected stock in your possession. 2. Clearly label the identified stock as "WITHDRAWN PRODUCT" DO NOT DISTRIBUTE OR USE. 3. Suspend any further distribution or transfer of the affected lots. 4. Provide written confirmation of the quantity and storage location of all affected stock within five (5) business days of receipt of this notification. 5. Our Quality team will coordinate collection, return, or replacement of affected product as appropriate." For questions or require further clarification, contact: Malcolm Applewhite, malcolm@glycar.co.za On 10/17/2025, the firm issued an URGENT: MEDICAL DEVICE RECALL notice the medical facility via email. This notice explained the issue, potential risk to health, and requested the following: "Actions to be Taken by the Customer / User: 1. Identify and quarantine any affected products with lot number T2502257 in your possession. 2. Cease use of the affected device immediately. 3. Share this notice with applicable personnel within your institution. 4. Abbott will coordinate and arrange for return and replacement of affected units. 5. Complete and return the attached Acknowledgment Form within 5 business days to confirm receipt of this notice." For questions or assistance, contact: Regulatory Affairs Manager, Glycar SA (Pty) Ltd, Malcolm Applewhite, malcolm@glycar.co.za
Root cause
Process control
Status
Open, Classified
Product code
DXZ
Quantity affected
18 units
Distribution
US
Date initiated
2025-10-09
Date posted
2025-11-25
Location
Irene, South Africa

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
GLYCAR PERICARDIAL PATCH (K963967)Glycar, Inc.1997-10-31