Atlas FDA

Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-

FDA device recall Z-1723-2009 · posted 2009-08-25 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. Two injuries and four complaints filed. 2/20/09: Recall being expanded to incl
Action taken
GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Healthcare sales or service representative.
Root cause
Device Design
Status
Terminated
Product code
LNH
Quantity affected
170(156 US / 14 OUS)
Distribution
Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, W
Date initiated
2008-11-03
Date posted
2009-08-25
Date terminated
2012-05-24