Atlas FDA

SIGNA MR355, SIGNA MR360, NMRI system

FDA device recall Z-1242-2025 · posted 2025-02-28 · Open, Classified

Recall details

Recalling firm
GE Healthcare (China) Co., Ltd.
Reason for recall
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Action taken
On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Root cause
Software design
Status
Open, Classified
Product code
LNH
Quantity affected
241
Distribution
Domestic US distribution nationwide. International distribution worldwide.
Date initiated
2025-01-31
Date posted
2025-02-28
Location
Beijing, China

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SIGNA Creator, SIGNA Explorer (K143251)Ge Healthcare (Ge Medical Systems, LLC)2015-02-04