Siemens Ysio Systems Product Usage: The Ysio (New RAD-FAMILY) systems are radiographic systems used in hospitals, clinic

FDA device recall Z-1717-2014 · posted 2014-06-04 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Excessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation.
Action taken
Siemens sent an Safety Advisory Notice letter dated May 9, 2014, to affected customer. The letter identified the affected product, problem and actions to be taken. For question call +49 (913) 84-9908.
Root cause
Use error
Status
Terminated
Product code
KPR
Quantity affected
433
Distribution
US Nationwide
Date initiated
2014-05-09
Date posted
2014-06-04
Date terminated
2015-08-18
Location
Malvern, PA