FDA device recall Z-2787-2011 · posted 2011-07-11 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
There is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the table may collide with extremities and cause injuries.
Action taken
Siemens sent a Safety Advisory Notice to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to place the Safety Notice in the System's instruction for use. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner. Customers were instructed to complete the Acknowledgement of Receipt Form and return to the firm.
Root cause
Other
Status
Terminated
Product code
MQB
Quantity affected
141
Distribution
Nationwide Distribution including Washington, DC and the following States: AR, CA, CO, FL, ID, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, a
Date initiated
2011-04-28
Date posted
2011-07-11
Date terminated
2014-04-10
Location
Malvern, PA