Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT O

FDA device recall Z-2180-2014 · posted 2014-08-10 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the inability to unload patient data.
Action taken
In June, 2014, an urgent medical device correction field safety notice of improvement was sent to end users informing them of the product, problem, and action to be taken to avoid the potential safety issue.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
56
Distribution
UT, WI, NY
Date initiated
2014-06-17
Date posted
2014-08-10
Date terminated
2014-12-15
Location
Malvern, PA