Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing modul
FDA device recall Z-2173-2012 · posted 2012-08-09 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sys
- Action taken
- Siemens Healtcare Diagnostics sent a Urgent Field Safety Notice dated June 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised to discuss the contents of this letter with your Laboratory Director regarding the need to review any previous test results. If you have any technical questions or concerns, please contact the Seimens Technical Solutions Center at 1-800-441-9250, for further assistance.
- Root cause
- Software design (manufacturing process)
- Status
- Terminated
- Product code
- NSX
- Quantity affected
- 15
- Distribution
- USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX.
- Date initiated
- 2012-06-07
- Date posted
- 2012-08-09
- Date terminated
- 2014-04-01
- Location
- Newark, DE