Atlas FDA

Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing modul

FDA device recall Z-2173-2012 · posted 2012-08-09 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sys
Action taken
Siemens Healtcare Diagnostics sent a Urgent Field Safety Notice dated June 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised to discuss the contents of this letter with your Laboratory Director regarding the need to review any previous test results. If you have any technical questions or concerns, please contact the Seimens Technical Solutions Center at 1-800-441-9250, for further assistance.
Root cause
Software design (manufacturing process)
Status
Terminated
Product code
NSX
Quantity affected
15
Distribution
USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX.
Date initiated
2012-06-07
Date posted
2012-08-09
Date terminated
2014-04-01
Location
Newark, DE