FDA device recall Z-0106-04 · posted 2003-11-13 · Terminated

Recall details

Recalling firm
Cti Pet Systems Inc
Reason for recall
CPS has found that in some instances, when the CT E-Stop is activated, a malfunction may occur where the PHS controller may not shut down and stop bed motion.
Action taken
Written notification was sent out beginning 10/02/2003 to distributors and end users to advise them of the problem.
Root cause
Other
Status
Terminated
Product code
KPS
Quantity affected
73
Distribution
Nationwide International: Japan, Germany, Brazil, Spain, China, Russia, Austrailia, Hong Kong, Singapore, Poland France, Korea, Greece, UK and Austria
Date initiated
2003-10-02
Date posted
2003-11-13
Date terminated
2006-02-07