Siemens Mevatron Linear Accelerator Systems: KD2, K 7767, K 8067, KD, KDS and KDS2, Mevatron KD2, Part No.: 1940753; Mev
FDA device recall Z-0851-2008 · posted 2008-02-19 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- Door Hinge failures; loose or falling off (Stationary Structure Doors)
- Action taken
- Stationary Door Hinges will be replaced on each device by the recalling firm's technician visits.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 1,284 total for recalls Z-0850-0852-2008
- Distribution
- Worldwide- USA including the states of Missouri, Georgia, Delaware, Texas, Ohio, Alabama, Kentucky, Florida, California, North Carolina, New Jersey, New York, Illinois, Pennsylvania, Colorado, Tenness
- Date initiated
- 2007-07-27
- Date posted
- 2008-02-19
- Date terminated
- 2011-04-29
- Location
- Concord, CA